ESIC circular · 02 Jun 2026
U-16/30/130/2012-Procell/E-792399 18-09-2026 To All Deans - ESIC PGIMSRs, Medical & Dental Colleges and Hospitals All Medical Superintendents - ESIC Hospitals Director (Medical) Delhi/Director (Medical) Noida Subject: Corrigendum in respect of Revised Specifications and cost estimation of various Equipment uploaded on E…
Official record
Open source page[Image omitted. See the official document.]
U-16/30/130/2012-Procell/E-792399
18-09-2026
To
Subject: Corrigendum in respect of Revised Specifications and cost estimation of various Equipment uploaded on ESIC Website for compliance by User Institutions– reg.
Ref: Hqrs Circular U-16/30/130/2012-Procell/E-792399, Dated: 02-06-2026
Attention is invited to this office letter circulated vide No. 16/30/130/2012-Procell/E-792399, Dated: 02-06-2026 (Copy attached) wherein Revised Specifications and Cost estimation of various equipment were issued for utilization by user institutions.
In this regard, it is informed that, the duly constituted Technical Specification Committee has reviewed and recommended certain modifications/changes in the existing technical specifications of the following equipment:
| S.No | Existing Specifications | Amended Specification |
|---|---|---|
| B10 | Additional Clause B (10) | System should have Shear Wave Elastography in Convex and Linear probes |
| C8 | Breast Elastography: Yes | Breast Elastography: Yes (Shear wave elastography) |
| C9 | Thyroid Elastography: Yes | Thyroid Elastography: Yes (Shear wave elastography) |
| D14 | Monitor display size, cm (in): 23" or more, preferably LED with handle | Monitor display size, cm (in): 23" or more, preferably LED |
| F2 | Linear probe should have automated quantification for easier identification of breast & Thyroid lesions (preferably with BIRAD & TIRAD based lesion classification tools | Linear probe should have Automated or semi-automated lesion quantification and characterization tools for breast and thyroid imaging, or should be upgradable to the same, preferably with BIRAD & TIRAD based lesion classification tools |
| F3 | System should have extensive obstetric measurement package including software for automatic measurement of NT | System should have extensive obstetric measurement package including so ware for automatic / assisted measurement of NT |
| F4 | System should have automatic tool to derive 9 planes of fetal heart (as per AIUM recommendation) from one volume sweep | System should provide tools for acquisition, reconstruction, and evaluation of fetal cardiac volume datasets, with automated and/or assisted methods for obtaining standard fetal cardiac views. |
| F5 | System should have automatic tool for deriving 9 planes for Fetal central Nervous System | System should provide automated and/or assisted tools for acquisition, reconstruction, and evaluation of standard fetal central nervous system imaging planes. |
| F11 | Quoted USG Model should be US FDA/CE/BIS approved. | Quoted USG Model should be US FDA / CE / BIS / ISO: 13485 approved |
| S.No | Existing Specifications | Amended Specification |
|---|---|---|
| Sl. No. G (Others) Bullet Point No. 2 | Emergency OFF switches mounted on monitor. | Emergency OFF switches should be provided |
| Sl. No. G (Others) Bullet Point No. 3 | Machine ON/OFF Key switch | Machine ON/OFF switch |
| Sl. No. G (Others) Bullet Point No. 4 | Flouro, Cine & spot switches on both side of panel | Flouro, Cine & spot switches should be provided |
| S.No | Existing Specifications | Amended Specification |
|---|---|---|
| Sl. No. I | High Frequency Generator 50KW generator should be latest technology with high frequency offering 50KHz or more. Constant power output of 40KW or more Generator should offer 3 point technique and user should have following parameters selection range. |
KV range should be 40 to 125
mA output: 500mA at100 KVP
mAs range should be 1 to 500mAs
Exposure time should be 1ms to 4 secs or better. | High Frequency Generator 50KW generator should be latest technology with high frequency offering 50KHz or more. User should have following parameters selection range.
KV range should be 40 to 125
mA output: 500mA at 100 KVP
mAs range should be 1 to 500mAs
Exposure time should be 1ms to 4 secs or better. | | Sl. No. II (1) | A Dual focus rotating anode X-ray tube with small focus 0.6 mm or less and large focus 1.2 mm or less | A Dual focus rotating anode X-ray tube with small focus 0.6 mm or less and large focus 1.5 mm or less | | Sl. No. II (3) | Tube rotation speed 9000 RPM or more. | Tube Rotation Speed: 2700 RPM or more | | Sl. No. III (1) | Floor to ceiling tube stand with counter balanced tube head | Floor to ceiling tube stand with counter balanced tube head or Ceiling Free tube stand | | Sl. No. IV (11) | Table accessories like stainless steel cassette tray, compression band should be provided. | Table accessories - compression band should be provided. | | Sl. No. IV (12) | Bucky tray should be equipped with cassette type wired or wireless flat panel detector. | Bucky tray should be equipped with wired or wireless flat panel detector. | | Sl. No. V (2) | Wall stand should have minimum of 3 field AEC | Vertical Bucky shall be equipped with Automatic Exposure Control (AEC) having minimum three selectable exposure sensing fields or equivalent detector-based exposure control technology | | Sl. No. VI (4) | The detector should be water resistant. | The detector should be water and dust resistant. | | Sl. No. VI (5) | Wireless detector if provided should be free from any integration with x ray machine. The same should be able to use with any mobile x-ray machine so that Detector can be used portably. The detector should be capable of doing out of bucky radiography in wireless mode and also lateral supine radiography must be possible | Wireless detector if provided should be free from any integration with x ray machine. Detector should support any xray system and should be capable of integrating with other x-ray machines. The detector should be capable of doing out of bucky radiography in wireless mode and also lateral supine radiography must be possible. | | Sl. No. VI (11) | The detector offered should be light in weight and less than 4 Kgs | The detector offered should be light in weight and less than 4.5 Kgs | | Sl. No. VI (13) | The battery must be latest Lithium ion type. 5 Nos batteries along with battery charger should be provided in case of wireless detector being offered. Fully charged battery should last for at least 5 hours in operating mode and 8 hours in energy saving mode. Company should replace batteries free of cost for next five years. The battery offered should be replaced irrespective of charge cycle limitations. In case of inbuilt battery detector manufacturer should either replace battery or detector, in case of any breakdown of battery. Kindly specify battery type and life of offered model. | The battery must be latest Lithium ion type. Total 5 Nos batteries along with battery charger should be provided in case of wireless detector being offered. Fully charged battery should last for at least 5 hours in operating mode and 8 hours in energy saving mode. Company should replace batteries free of cost for next five years. The battery offered should be replaced irrespective of charge cycle limitations. In case of inbuilt battery detector manufacturer should either replace battery or detector, in case of any breakdown of battery. Kindly specify battery type and life of offered model. | | Sl. No. VI (14) | Detector offered should be capable of integrating with any x-ray system or mobile x-ray and should be readily switchable within multiple x- ray machine, in case there is breakdown in the x-ray system | Detector should support any xray system and should be capable of integrating with other x-ray machines. | | Sl. No. VII (3) | Tubal focal spot selection switch. | Tubal focal spot selection switch (Manual/Automatic) | | Sl. No. VIII (25) | DR console should be connected to a compatible dry laser printer with resolution 500 DPI or more | DR console should be connected to a compatible dry laser printer provided by the vendor having resolution 500 DPI or more | | Sl. No. X (15) | DAP meter | DAP meter (either inbuilt or as accessories) | | Sl. No. XI (3) | Company should have ISO 9001:2015 and ICMED 13485 certificate. Also have MD 9 License for Medical Devices from CDSCO for quoted model | Company should have ISO 9001:2015, ISO-13485 / ICMED13485 Certificate and MD 9 License for Medical Devices from CDSCO for quoted model. | | Sl. No. XI (9) | The bidder and the principal company should have a good reputation and never been black - listed or debarred in any state and central government organization | Deleted | | General Points | All software upgrades should be done free of cost by the OEM/bidder within one month of issue of new upgrade from Principal OEM. | All software update should be done free of cost by the OEM/bidder within one month of issue of new upgrade from Principal OEM within warranty and CMC period. |